Name of measuring instrument Parameter Category/ Typical model Name, Code of calibration regulation Measuring range Best Measurement Capability Expanded As An uncertainty
15 Optical Power Meter for Telecommunication Range of Optical Power HP 8156A, HP 8158A, OLA-100, OLA-15 JJG 9652001 Verification Regulation of Optical power Meter in Telecommunication 0~50dBm U=1%(k=2)
Value of Optical Power
16 Optical Attenuator for optical communication Value of Attenuator HP 8156A, HP 8158A, OLA-100, OLA-15 JJG (YD) 04098 Verification Regulation of Optical Attenuator (10 ~ 60) dB ≤0.2%(10~40dB)             ≤0.5% ( >40dB )
Return Loss of Attenuator (Option)
17 Optical Time Domain Reflectometer (OTDR) Distance accuracy HP8147 HPE6000 PK7500 JJG 959-2001 Verification Regulation of Optical Time Domain Reflectometer (0-300) km 0.2m
Loss measurement (0-40)dB 0.016dB/dB
18 Stabilized Laser Sources for Optical Transmit Output Power OLS-100, HP81554SM JJG 9582000 Verification Regulation of Stabilized Laser Sources for Optical Transmit 10mW~ 1pW (2~ 10 )%  Stability: ≤.05dB/ 15min  ≤0.3 dB/1h
Stability of output power
Spectrum
19 Optical Wavelength Optical Wavelength Optical Telecommunications MF 9630A HP86120x etcs JJG963-2001 Optical Wavelength Counters in Telecommunication (0.6~1.6)mm U=5´10-7(k=2)
20 Optical Spectrum Analyzer Optical Wavelength Optical Telecommunications MS9710,HP71518614X,etcs JJG(YD)046-99 Verification Regulation of the Optical Spectrum Analyzer (0.6~1.6)mm U=5´10-7(k=2)
Optical Power (+3~-110)dBm £0.4dB
Resolution Bandwidth   £0.05nm
Dynamic Range   £70dB(1nm)
28 Tunable Laser Source
Wavelength range
1.1.          Tunable Laser/ 1.2.          MS 9638A 1.3.            TUNICS-PRI、 1.4.          Agilent 1.5.          81640A 1.6.          81689A 1.7.           etc JJG(YD)055-2002 Verification Regulation of Tunable Laser Source (600~1600)nm

U=0.005nm (k=2)
Wavelength accuracy U=5´10-7k=2)
Wavelength resolution U=0.0005nm (k=2)
Wavelength repeatability U=0.0005nm (k=2)
Wavelength stability U=0.0005nm (k=2)
Power stability (+3~-110)dBm U=0.005dB (k=2)
Output Power (+30~-50)dBm (600~1750)nm U=3% (k=2)
Side-mode suppression ratio U=0.3dB (k=2)
Line width (1200~1600)nm 100kHz~22GHz U=0.05dB (k=2)
29 Chromatic Dispersion Analyzer for Single Mode Optical Fiber Zero Dispersion Wavelength S18 CD400 JJG(YD)0060-2002 Chromatic Dispersion Analyzer for Single Mode Optical Fiber 1.8.            (1310 ±40)nm, 1.9.          (1550±40)nm 1.10.        0.5nm
Slope at Zero Dispersion Wavelength 0.005ps/nm*km

 

 

 